Desktop Conductivity Meter EZL-DCM123 www.Ezilab.com |
[email protected]
Overview The Desktop Conductivity Meter EZL-DCM123 comes with a waterresistant membrane keyboard and a big dual display as standard. The conductivity measuring range of the model is 0.00 µS/cm to 2000 mS/cm. Temperature variations are controlled between 5 °C and 35 °C, with a TDS of 0 to 100 g/L, a resistivity of 0 to 100 MW.cm, and an automatic temperature compensation ranging from 0 °C to 100 °C with an accuracy of 5 °C to 60 °C: 0.3 °C others: 0.5 °C.
Specifications :
Measuring Range
0.00 to 20.00 µS/cm 20.0 to 200.0 µS/cm 200 to 2000 µS/cm 2.00 to 20.00 mS/cm 20.0 to 200.0 mS/cm 200 to 2000 mS/cm
Resistivity
0 to 100 MΩ.cm
TDS
0 to 100 g/L
Salinity
0 to 100 ppt
Resolution
0.01/0.1/1 µS/cm 0.01/0.1 mS/cm Electrode: ± 0.5 % FS
Accuracy Instrument: ± 1.00 % FS Temperature Compensation
0 °C to 100 °C (Automatic)
Electrode Constant
0.1/1/10 cm-1
Temperature Measuring Range
10 °C to 110 °C
Temperature Resolution
0.1 °C
Temperature Accuracy
5 °C to 60 °C: ± 0.3 °C Others: ± 0.5 °C
Ambient Temperature
5 °C to 35 °C
Ambient Humidity
≤ 80 %
Power
DC 12V/1A
Dimensions (W×D×H)
240×170×70 mm
Weight
600 g
Features : - 7-inch high resolution capacitive touch screen - Auto-calibration, automatic temperature compensation electrode, data storage, USB port, self-diagnosis functions - Bilingual operation - Data storage of 1000 sets of measurement data and is transferable - Auto-recognition of 8 conductivity standard solutions - Bluetooth module and wireless printer - Optional PC software and cell phone APP - 1 point calibration of electrode meet accuracy for 0.5µS/cm – 200mS/cm - Switch between conductivity, TDS, salinity and resistivity - Multinomial calculation of TDS, salinity - Smart judge electrode status for accurate results - Dustproof and waterproof meter meet IP54 Grade - Has a “ready” indicator and HOLD function
Applications : It has a wide range of applications, including high precision conductivity measurements in scientific research, quality control, biotechnology, and fine chemistry, such as measuring the conductivity of common water solutions, high purity water, reverse osmosis water, and distilled water.
Standard Configurations : Desktop Conductivity Meter LO-DCM100 Series 1 pc Composite Conductivity Electrode 1 pc Power Adapter 1 pc Electrode Holder and Base 1 set Instruction Manual in USB 1 copy
Ezilab Scientific Email:
[email protected] | Website: www.Ezilab.com